inforMED
MalfunctionJTO

VITEK® 2 NEISSERIA-HAEMOPHILUS IDENTIFICATION TEST KIT

Received Dec 15, 2016

Report 1950204-2016-00205 · MDR key 6177522

Device

Generic name

Vitek® 2 Nh Id Test Kit

Manufacturer

Biomerieux Inc.

Catalog number

21346

Lot number

245369522

Product problems

  • Incorrect Or Inadequate Test Results
  • Incorrect Or Inadequate Test Results

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

A CUSTOMER REPORTED A MISIDENTIFICATION OF NEISSERIA MENINGITIDIS AS NEISSERIA GONORRHOEAE BY THE VITEK® 2 NH TEST KIT LOT # 245369522 WITH VITEK 2 7.01 SOFTWARE. THE IDENTIFICATION TO NEISSERIA MENINGITIDIS WAS CONFIRMED THROUGH PCR AND MS. THE CUSTOMER DID NOT SUBMIT STRAIN OR RAW DATA FOR ANALYSIS. AN INVESTIGATION WAS CONDUCTED. THE CUSTOMER HAD SET UP THE PATIENT ISOLATE FROM A 48-HOUR CULTURE GROWN ON CHOCOLATE AGAR. USE OF A 48-HOUR CULTURE IS NOT RECOMMENDED. THE AGE OF CULTURE SHOULD BE 18-24 HRS. A LAB REPORT WAS SUBMITTED THAT SHOWED A VERY GOOD IDENTIFICATION OF NEISSERIA GONORRHOEAE WITH FIVE (5) ATYPICAL POSITIVE REACTIONS (ARGA, LYSA, TYRA, PVATE, APPA), AND ONE (1) ATYPICAL NEGATIVE REACTION (DMAL), FOR AN IDENTIFICATION OF NEISSERIA MENINGITIDIS ACCORDING TO THE VITEK® 2 NH KNOWLEDGE BASE. AN INCREASED NUMBER OF ATYPICAL POSITIVE REACTIONS CAN INDICATE CONTAMINATION, MIXED CULTURE, USE OF NON RECOMMENDED MEDIA OR OTHER USER SET UP ERRORS OR AN ATYPICAL STRAIN. HOWEVER WITHOUT THE STRAIN OR RAW DATA IT'S NOT POSSIBLE TO FURTHER EVALUATE THE CAUSE OF THE MISIDENTIFICATION. A REVIEW OF QUALITY RECORDS CONFIRMED THE NH ID LOT 245369522 MET QC CRITERIA FOR INITIAL PERFORMANCE TESTING AND FINAL RELEASE. IN CONCLUSION, THE CUSTOMER TESTED THE ISOLATE FROM 48-HOUR CULTURE, RATHER THAN 18-24 HOURS AS SPECIFIED IN THE NH ID PRODUCT INFORMATION. THE CUSTOMER DID NOT SUBMIT THE ISOLATE OR RAW CARD DATA, THEREFORE THE MISIDENTIFICATION COULD NOT BE CONFIRMED.

Description of Event or Problem

A CUSTOMER FROM THE UNITED STATES REPORTED TO BIOMÉRIEUX A DISCREPANT RESULT FOR A URINE SAMPLE IN ASSOCIATION WITH THE VITEK® 2 NH TEST KIT. THE CUSTOMER INITIALLY TESTED THE SAMPLE USING THE VITEK® MS WHICH IDENTIFIED NEISSERIA MENINGITIDES. CONFIRMATION TESTING OF THE ISOLATE WITH VITEK® 2 NH GAVE A RESULT OF (B)(6). A REPEAT TEST WITH THE VITEK® MS IDENTIFIED NEISSERIA MENINGITIDES AND PCR WAS (B)(6). THE RESULT OF NEISSERIA MENINGITIDES WAS REPORTED TO THE PHYSICIAN. THERE IS NO INDICATION OR REPORT FROM THE HOSPITAL OR TREATING PHYSICIAN TO BIOMÉRIEUX THAT THE DISCREPANT RESULT LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. THE CUSTOMER DID NOT REPEAT THE TEST FOR VITEK® 2 NH AS THEY DID NOT HAVE ENOUGH ISOLATE COLONIES. THE CUSTOMER DOES NOT HAVE THE ISOLATE FOR EVALUATION. AN INTERNAL BIOMÉRIEUX INVESTIGATION HAS BEEN INITIATED.