CLINITEK ADVANTUS
Received Dec 9, 2016 · Event occurred Nov 11, 2016
Report 3002637618-2016-00165 · MDR key 6157805
Device
Generic name
Clinitek Advantus
Catalog number
10636672
Product problems
- Use of Device Problem
- Use of Device Problem
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
THE CUSTOMER CONFIRMED THAT THE OPERATOR NOTICED THE DISCREPANCY IN A NUMBER OF RESULTS OF 6 PATIENTS THAT WENT ACROSS THE INTERFACE AND IMMEDIATELY ISSUED CORRECTED REPORTS AND STATED THAT NO PATIENT HARM OCCURRED DUE TO THIS ISSUE. THE CUSTOMER REPORTED THAT A REPLACEMENT BARCODE READER WAS RECEIVED.
Description of Event or Problem
THE CUSTOMER REPORTED THAT THE BARCODE READER ON THEIR SYSTEM MISREAD THE PATIENT BARCODE LABEL RESULTING IN WRONG RESULTS BEING SENT ACROSS LIS. THERE WAS NO REPORT OF INJURY DUE TO THIS EVENT.
Additional Manufacturer Narrative
(B)(4).