inforMED
MalfunctionCIT

AST IFCC WITHOUT P-5-P

Received Aug 25, 2004 · Event occurred Aug 10, 2004

Report 1823260-2004-00767 · MDR key 612425

Device

Generic name

Clinical Chemistry Reagent

Manufacturer

Roche Diagnostics

Model number

NA

Catalog number

11928511216

Lot number

ALL

Product problems

  • Unknown (for use when the device problem is not known)

Patient

NA

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

INCORRECT LOW AND ZERO PATIENT RESULTS MAY OCCUR WITH THIS REAGENT ON THE HITACHI MODULAR P ANALYZER. ROTATION OF THE INSTRUMENT REAGENT COMPARTMENT CAN CAUSE A BUBBLE TO FORM IN THE NECK OF THE REAGENT BOTTLE CAUSING PREMATURE LIQUID LEVEL DETECTION AND SHORT SAMPLING OF THE REAGENT. THIS ISSUE APPEARS TO BE RELATED TO FILL VOLUME, VISCOSITY, AND BOTTLE NECK DESIGN.

Remedial action

  • Recall