inforMED
InjuryFHM

PREMIUM MF TA* 30-V3 TITANIUM STAPLER

Received Nov 11, 2016

Report 1219930-2016-01182 · MDR key 6096464

Device

Generic name

Apparatus, Suturing, Stomach And Intestinal

Model number

010315

Catalog number

010315

Lot number

N4F0438X

Product problems

  • Positioning Failure
  • Positioning Failure

Patient

Not reported

  • Blood Loss
  • Blood Loss

Narrative

Additional Manufacturer Narrative

(B)(4). POST MARKET VIGILANCE (PMV) LED AN EVALUATION OF ONE PREMIUM MF TA 30-V3 TITANIUM STAPLER AND TWO PREMIUM MF TA 30-V3 TITANIUM DLU . THIS EVALUATION WAS BASED ON A MEDICAL AND TECHNICAL REVIEW OF ALL DATA RECEIVED FROM THE SITE, A PMV REVIEW OF MANUFACTURING RECORDS, A PMV REVIEW OF COMPLAINT TRENDS AND AN EVALUATION OF THE RETURNED DEVICE. THE JAWS WERE OPEN. THE SULUS WERE FULLY FIRED; ONE SULU HAD TWO MALFORMED STAPLE. THE SULUS WERE RELOADED WITH STAPLES. THE INSTRUMENT AND SULUS WERE FIRED OVER TEST MEDIA WITH ACCEPTABLE STAPLE FORMATION. SHOULD NEW INFORMATION BECOME AVAILABLE, THE FILE WILL BE RE-OPENED AND THE INVESTIGATION SUMMARY WILL BE AMENDED AS APPROPRIATE.

Additional Manufacturer Narrative

(B)(6).

Description of Event or Problem

ACCORDING TO THE REPORTER: DURING A THORACIC PROCEDURE, THE DOCTOR ATTEMPTED TO FIRE THE DEVICE ON 2 OCCASIONS RESULTING IN BLEEDING. STAPLER DEPLOYMENT WAS INCOMPLETE, RESULTING IN ONLY PARTIAL LIGATION OF THE PULMONARY VEIN. THE DOCTOR SUBSEQUENTLY LIGATED THE REMAINDER OF THE PULMONARY VEIN WITH PROLENE SUTURE TO ACHIEVE HEMOSTASIS. THE PATIENT WAS TRANSFERRED TO PACU IN STABLE CONDITION AFTER SURGERY.