inforMED
MalfunctionGET

AIR HOSE 5.0M AESC.-DRAEGER/AESC.LARGE

Received Oct 27, 2016

Report 2916714-2016-00936 · MDR key 6062021

Device

Generic name

High Speed Power Systems

Manufacturer

Aesculap Ag

Model number

GA466R

Catalog number

GA466R

Lot number

51977098

Product problems

  • Burst Container or Vessel
  • Burst Container or Vessel

Patient

Not reported

  • Injury
  • Injury

Narrative

Additional Manufacturer Narrative

INVESTIGATION: THE INVESTIGATION WAS CARRIED OUT BY THE AESCULAP TECHNICAL SERVICE DEPARTMENT (ATS). IT HAS BEEN ASCERTAINED, THAT AN OLD VERSION OF THE HOSE WAS USED. THIS HOSE WAS PRONE TO FLAWS WHICH COULD LEAD TO A BURST. THEREFORE THIS HOSE HAS BEEN CHANGED VIA THE INTERNAL CHANGE REQUEST: (B)(4). BATCH HISTORY REVIEW: THE DEVICE HISTORY FILES HAVE BEEN CHECKED FOR THE AVAILABLE LOT NUMBER AND FOUND TO BE ACCORDING TO THE SPECIFICATION, VALID AT THE TIME OF PRODUCTION. CONCLUSION AND ROOT CAUSE: THIS FAILURE OCCURRED MOST LIKELY DUE TO DESIGN ERROR. CORRECTIVE ACTION: A CHANGE HAS BEEN INITIATED TO ELIMINATE THE FAILURE, SEE CHANGE. (B)(4).

Additional Manufacturer Narrative

(B)(4) (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF (B)(4) (MANUFACTURER). EXEMPTION NUMBER: E2014018. G4. IT SHOULD BE NOTED THAT COMPLAINT INFORMATION WAS RECEIVED 07/07/2016; INFORMATION REGARDING BLAST INJURY WAS NOT RECEIVED UNTIL 10/24/2016. MANUFACTURING SITE EVALUATION: EVALUATION ON-GOING.

Description of Event or Problem

COUNTRY OF COMPLAINT: GERMANY. IT WAS REPORTED THAT 5 PEOPLE FROM THE OPERATION TEAM WERE INJURED WHEN THE AIR HOSE BURST.