TEMPOROMANDIBULAR JOINT-PATIENT MATCHED IMPLANT
Received Oct 19, 2016 · Event occurred Sep 20, 2016
Report 0001032347-2016-00602 · MDR key 6041877
Device
Generic name
Patient Matched Implant
Manufacturer
Biomet MicrofixationModel number
N/ACatalog number
TMJPMI-1532
Lot number
707830A
Product problems
- Component Missing
- Component Missing
Patient
61 YR
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
(B)(4). A REVIEW OF THE DEVICE HISTORY RECORDS SHOW THE LOT RELEASED WITH NO RECORDED ANOMALY OR DEVIATION. THE USER FACILITY IS FOREIGN; THEREFORE A FACILITY MEDWATCH REPORT WILL NOT BE AVAILABLE. THE PRODUCT WAS IMPLANTED INTO THE PATIENT AND THEREFORE WILL NOT BE RETURNED FOR AN EVALUATION. CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A VALID CONCLUSION ABOUT THE CAUSE OF THIS EVENT. IF ADDITIONAL INFORMATION IS OBTAINED THAT ADDS VALUE TO THE RELEVANT CONTENT OF THIS REPORT AND/OR A CONCLUSION CAN BE DRAWN, A FOLLOW-UP REPORT WILL BE SENT.
Description of Event or Problem
IT WAS REPORTED THAT A TEMPLATE AND A FOSSA IMPLANT FROM A TEMPORMANDIBULAR JOINT (TMJ) PATIENT MATCHED IMPLANT (PMI) SYSTEM WERE MISSING A HOLE. IT IS REPORTED THAT THE SURGEON CREATED THE MISSING HOLES. IT WAS REPORTED THAT THIS EVENT PRESENTED NO DELAY TO THE PROCEDURE AND NO INJURY TO THE PATIENT.
Additional Manufacturer Narrative
THE COMPLAINT IS THE SUTURE HOLE IS MISSING FROM THE FOSSA TEMPLATE AND IMPLANT. THE COMPLAINT WAS CONFIRMED THROUGH INTRAOPERATIVE PICTURES. THE MOST LIKELY UNDERLYING CAUSE IS THE DESIGN VENDOR SENDING THE INCORRECT REVISION OF THE CAD (COMPUTER AIDED DESIGN) FILE TO ZIMMER BIOMET FOR MANUFACTURE.