inforMED
MalfunctionGIO

CAPILLARY ADAPTERS

Received May 16, 2005 · Event occurred Mar 4, 2005

Report 3002807968-2005-00004 · MDR key 603669

Device

Generic name

Adaptor For Connecting Capillary Tubes

Model number

906-018

Catalog number

*

Lot number

ALL 2005

Product problems

  • Bubble(s)

Patient

*

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

A SMALL PLASTIC ADAPTER IS NECESSARY IN ORDER TO TRANSFER A BLOOD SAMPLE FROM A CAPILLARY TUBE TO A BLOOD GAS ANALYZER TYPE ABL5. FOLLOWING A TRANSFER THE USER NOTICED THAT THE PO2 VALUES WERE TOO HIGH AND APPROACHING THAT OF AMBIENT AIR. A BUBBLE WAS SUBSEQUENTLY DISCOVERED IN THE MEASURING CHAMBER. OTHE TESTS REVEALED THAT THE INTRODUCTION OF AMBIENT AIR IS POSSIBLE WHEN USING THE ADAPTER. NO PT WAS MALTREATED.