inforMED
InjuryMOS

NOT DETERMINED

Received Sep 29, 2016 · Event occurred Aug 29, 2016

Report 1056069-2016-00003 · MDR key 5988590

Device

Generic name

Not Determined

Manufacturer

Invivo Corporation

Lot number

N/A

Product problems

  • Insufficient Device Problem Information

Patient

5 MO

  • Partial thickness (Second Degree) Burn
  • Partial thickness (Second Degree) Burn

Narrative

Additional Manufacturer Narrative

BASED ON THE PROVIDED INFORMATION AND TEST PERFORMED ON SITE, THERE IS NO INDICATION OF A MALFUNCTION OF THE COIL USED. NO DEFINITE CAUSE FOR THE INJURY ON THE PATIENT LEFT LEG COULD BE DETERMINED. THE COIL WAS CHECKED AFTER THE EXAMINATION. THE COIL PASSED THE PERFORMED TESTS. NO DAMAGE WAS OBSERVED AND THE COIL DID NOT BECOME TOO HOT TO TOUCH. THE SPO2 SENSOR WAS CLOSE TO THE AFFECTED AREA BUT IS NOT LIKELY TO HAVE CAUSED THE INJURY. THE SENSOR IS MR SAFE AND THE MONITORING DEVICE WITH THE SAME ACCESSORIES WAS USED BEFORE AND AFTER THE INCIDENT IN A SIMILAR SET UP WITHOUT PROBLEMS. THE PATIENT WAS COVERED BY A BLANKET OF 100% POLYESTER. THE BLANKET MAY HAVE CONTRIBUTED TO THE INCIDENT; HOWEVER, THE DEVELOPMENT OF THE BLISTER WAS NOT OBSERVED AS THE AREA WAS COVERED WITH THE BLANKET. ALSO, THE BLANKET IS MADE OF POLYESTER; POLYESTER IS AN INSULATOR AND THEREFORE HINDERS THE HEAT DISSIPATION. THE BLANKET DOES NOT CONTAIN CONDUCTIVE MATERIALS. NEITHER THE ELEPHANT APPLICATION, NOR THE SEAMS (YARN) ARE SUSPECTED TO BE CONDUCTIVE (JUDGED BY PHOTOGRAPHIC EVALUATION). THE NURSE TREATED THE AFFECTED SKIN AREA. THE SKIN IS EXPECTED TO HEAL COMPLETELY. THE CUSTOMER WAS INFORMED TO ALWAYS FOLLOW THE INSTRUCTIONS FOR USE OF EACH ACCESSORY USED DURING AN EXAMINATION. IT IS REQUIRED EVEN WITH MR CONDITIONAL /SAFE EQUIPMENT TO PREVENT DIRECT CONTACT BETWEEN CABLING AND THE PATIENT'S BODY. BASED ON THE PROVIDED INFORMATION AND TEST PERFORMED ON SITE, THERE IS NO INDICATION OF A MALFUNCTION OF THE COIL USED.

Additional Manufacturer Narrative

EVENT IS UNDER INVESTIGATION AND UPDATE WILL BE PROVIDED ON A LATER DATE.

Description of Event or Problem

A SEDATED INFANT WAS SCANNED WITH THE NVC 1.5T COIL HEAD FIRST IN A SUPINE POSITION FOR A BRAIN SCAN WHEN THE PATIENT RECEIVED A 2ND DEGREE BURN ON THEIR FOOT APPROXIMATELY 3CM BLISTER.

Remedial action

  • Other