SORIN S5 BUBBLE DETECTOR FOR 3/8 TUBING
Received Sep 27, 2016 · Event occurred Oct 15, 2015
Report 9611109-2016-00678 · MDR key 5982188
Device
Generic name
Monitor And/or Control, Level Sensing, Cardiopulmonary Bypass
Manufacturer
Sorin Group DeutschlandModel number
23-07-50Product problems
- Physical Property Issue
- Physical Property Issue
Patient
Not reported
- No Patient Involvement
- No Patient Involvement
Narrative
Additional Manufacturer Narrative
SORIN GROUP (B)(4) MANUFACTURES THE SORIN S5 BUBBLE DETECTOR. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH REPORT IS BEING FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE SORIN S5 BUBBLE DETECTOR WAS FOUND TO HAVE DEFLATED CUSHIONS. THIS ISSUE WAS DISCOVERED BY A SORIN GROUP FIELD SERVICE REPRESENTATIVE DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT. THE SERVICE REPRESENTATIVE REPLACED THE BUBBLE SENSOR AND PERFORMED A FUNCTIONAL CHECK AND TRIAL RUN. NO FURTHER ISSUES WERE DISCOVERED. THE INVESTIGATION IS ONGOING. A FOLLOW-UP REPORT WILL BE SENT WHEN THE INVESTIGATION IS COMPLETE.
Description of Event or Problem
SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE SORIN S5 BUBBLE DETECTOR WAS FOUND TO HAVE DEFLATED CUSHIONS. THIS ISSUE WAS DISCOVERED BY A SORIN GROUP FIELD SERVICE REPRESENTATIVE DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.
Remedial action
- Other