UNKNOWN NEONATAL
Received Sep 23, 2016
Report 3009211636-2016-00410 · MDR key 5975428
Device
Generic name
Umbilical Vessel Catheter
Manufacturer
CovidienModel number
UNKNOWN NEONATALCatalog number
UNKNOWN NEONATAL
Product problems
- Material Fragmentation
Patient
Not reported
- Foreign Body Reaction
- Foreign Body Reaction
Narrative
Additional Manufacturer Narrative
AN INVESTIGATION IS CURRENTLY UNDERWAY. UPON COMPLETION, THE RESULTS WILL BE FORWARDED.
Description of Event or Problem
IT WAS REPORTED TO COVIDIEN ON (B)(6) 2016 THAT A CUSTOMER HAD AN ISSUE WITH AN UMBILICAL VESSEL CATHETER (UVC). THE CUSTOMER REPORTS A 3.5 DUAL LUMEN UMBILICAL VESSEL CATHETER (UVC) WAS PLACED EMERGENTLY IN THE MAIN OR FOLLOWING DELIVERY OF A 0.415 GRAM INFANT VIA EMERGENCY CAESARIAN SECTION. THE FLUSHED 3.5 FR DOUBLE LUMEN CATHETER WAS ADVANCED TO 7 CM WITHOUT DIFFICULTY. THE LINE FLUSHED AND DREW BACK EASILY, A 3.0 SILK SUTURE WAS NOT AVAILABLE IN THE OR ROOM AT THE TIME; HOWEVER, A 3.0 COATED SUTURE WAS AVAILABLE. THE CATHETER WAS SUTURED IN PLACE AT 7CM WITH THE 3.0 COATED SUTURE. AFTER TRANSFERRING THE INFANT TO THE NICU AN X-RAY WAS TAKEN OF THE LINE PLACEMENT AND SHOWED THE UVC NEEDED TO BE ADJUSTED BACK 1.5 CM. WHILE THE BEDSIDE NURSE WAS PULLING THE UVC BACK TO THE CORRECT POSITION, THE CATHETER SNAPPED AT THE 6CM MARK. THE NURSE WAS UNABLE TO GRAB THE PROXIMAL TIP AND THE CATHETER SUNK DEEPER INTO THE UMBILICAL STUMP, 6CM OF CATHETER REMAINED IN THE UMBILICAL VEIN AND WAS NOT ABLE TO BE RETRIEVED. THE PATIENT WAS TRANSFERRED TO A SPECIALTY HOSPITAL FOR SURGICAL REMOVAL OF THE CATHETER.