inforMED
InjuryMOS

DS ANTERIOR 1.5T

Received Jul 1, 2016

Report 1056069-2016-00001 · MDR key 5764841

Device

Generic name

Ds Anterior 1.5t

Manufacturer

Invivo Corporation

Model number

9896-030-23954

Lot number

N/A

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

59 YR

  • Burning Sensation
  • Partial thickness (Second Degree) Burn
  • Burning Sensation
  • Partial thickness (Second Degree) Burn

Narrative

Additional Manufacturer Narrative

EVENT IS UNDER INVESTIGATION AND UPDATES WILL BE PROVIDED ON A LATER DATE. (B)(4).

Description of Event or Problem

A MALE PATIENT WAS POSITIONED HEAD FIRST SUPINE AND SCANNED FOR PELVIS EXAMINATION USING THE DS ANTERIOR 1.5T COIL. THE PATIENT EXPERIENCED A HEATING SENSATION AND 2ND DEGREE BURNS RESULTING IN BLISTERS ( APPROX. 2CM X 2CM) ON THE INSIDE OF BOTH INNER THIGHS.

Additional Manufacturer Narrative

BASED ON THE PROVIDED INFORMATION AND TEST PERFORMED ON SITE, THERE IS NO INDICATION OF A MALFUNCTION OF THE COIL USED THAT CONTRIBUTED. THE INJURIES ON THE INSIDE OF THE PATIENT¿S INNER THIGHS ARE CONSISTENT WITH INJURIES CAUSED BY SO CALLED SKIN-TO-SKIN CONTACT. ALTHOUGH PADDING WAS USED BETWEEN THE LEGS THIS COULD NOT PREVENT THE UPPER INNER THIGHS FROM TOUCHING. IT WAS STATED THAT THERE WAS POTENTIAL FOR SKIN-TO-SKIN CONTACT CLOSE TO THE AFFECTED AREA. FURTHERMORE, 3 SCANS ON HIGH SAR LEVEL WERE ADMINISTERED THAT MAY HAVE CONTRIBUTED.

Remedial action

  • Other