inforMED
MalfunctionFEN

ABVISER AV IAP DEVICE, POLE MOUNT (1X10)

Received Jun 16, 2016 · Event occurred May 28, 2016

Report 1049092-2016-00282 · MDR key 5730246

Device

Generic name

Device, Cystometric, Hydraulic

Manufacturer

Convatec, Inc.

Model number

ABV301

Lot number

140600

Product problems

  • Disconnection
  • Disconnection

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

A BATCH RECORD REVIEW FOR LOT# 140600 WAS PERFORMED BY THE ORIGINAL EQUIPMENT MANUFACTURER (OEM) AND SHOWED THAT ALL FUNCTIONAL TESTING MET THE SPECIFICATION. THERE WERE TWO SUB-ASSEMBLY LOT#S; 140428 AND 140444 USED TO MANUFACTURE LOT# 140600. REVIEW OF LOT# 140428 SHOWS THAT ALL FUNCTIONAL TESTING MET SPECIFICATION. HOWEVER, LOT# 140444 DID HAVE A NON-CONFORMANCE RELATED TO THE QA IN-PROCESS FLOW TEST PROCESS. THE NON-CONFORMANCE DOES NOT AFFECT THE OVERALL INTACT OF THE VALVE (I.E.,THE NON-CONFORMANCE IS NOT RELATED O THE COMPLAINT ISSUE.) A PHYSICAL SAMPLE WAS RECEIVED FOR THIS COMPLAINT, AND EVALUATED. THE USED (DECONTAMINATED) PRODUCT SAMPLE WAS RETURNED TO THE OEM WHERE IT WAS EXAMINED UNDER A MICROSCOPE. THE PRODUCT SAMPLE SHOWED MAJOR CRACKS ALL AROUND THE VALVE CAP AND VALVE BODY AREA, WHICH COULD CAUSE THE VALVE TO DETACH DURING USAGE. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET PRODUCT MONITORING REVIEW PROCESS. NO ADDITIONAL PATIENT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative

RECEIVED (1) ONE USED ABVISER FOR EVALUATION. THE RETURNED PRODUCT SAMPLE HAS BEEN FORWARDED TO THE ORIGINAL EQUIPMENT MANUFACTURER FOR FURTHER INVESTIGATION. NO ADDITIONAL PATIENT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. (B)(4).

Additional Manufacturer Narrative

BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A MALFUNCTION. THE INSTRUCTIONS FOR USE SAY TO ENSURE THE URINARY CATHETER TO VALVE CONNECTION IS DRY AND THEN WRAP A STRIP OF TAPE (INCLUDED IN THE PACKAGE) AROUND THE END OF THE CATHETER TO KEEP IT SNUGLY ATTACHED TO THE ABVISER AUTOVALVE VALVE. NO ADDITIONAL PATIENT/EVENT DETAILS HAVE BEEN PROVIDED TO DATE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem

INFORMATION REPORTED BY A DISTRIBUTOR INDICATED THAT THE AUTO VALVE OF THE ABVISER DISCONNECTED FROM THE FOLEY CATHETER WHILE INJECTING NORMAL SALINE. THE DEVICE HAD BEEN IN USE FOR FORTY HOURS. THE INDICATION FOR USE OF THE DEVICE WAS INTESTINAL SWELLING WHICH NEEDED REGULAR INTRA-ABDOMINAL PRESSURE (IAP) MONITORING. THE PATIENT WAS IN THE SUPINE POSITION WITH THE HEAD OF THE BED FLAT WHEN THE MEASUREMENT WAS TAKEN,THERE WAS NO RESISTANCE FELT AT THE TIME OF THE EVENT. IT IS UNCLEAR IF THE BLUE TAPE WAS PROVIDED WITH THE KIT OR WAS APPLIED TO THE CONNECTION BETWEEN THE AUTO-VALVE AND THE CATHETER, AS THIS INFORMATION WAS UNAVAILABLE. THE ABVISER WAS REPLACED WITH A NEW ONE; NO HARM WAS REPORTED TO THE PATIENT AS A RESULT OF THE DISCONNECTION. THE PATIENT'S CURRENT STATUS IS BEING KEPT IN SURGICAL INTENSIVE CARE UNIT UNDER OBSERVATION.