ORS-100N SOLUTION WARMER DRAPE
Received Jun 16, 2016
Report 8043817-2016-00034 · MDR key 5729293
Device
Generic name
Fluid Warming And Slush Drapes
Manufacturer
Microtek Dominicana, S.a.Model number
ORS-100NLot number
SEE H10
Product problems
- Break
- Fluid/Blood Leak
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
THIS EVENT WAS ASSOCIATED WITH ITEM NUMBER OM-ORS-100, LOT NUMBER 1052213 AND IS ASSOCIATED WITH THE POTENTIAL FOLLOWING ORS-100N NON STERILE MICROTEK LOT NUMBERS, D153021, D151621A, D152941, D151971, D151751, D151611, D151341, D153081A, D152761A, D153081, D153071, D153091, D153011, D153011A, AND D152891. NO DEVICE WAS RETURNED FOR EVALUATION. BASED ON THE DEVICE HISTORY RECORD REVIEW, THE NON CONFORMITY DOES NOT APPEAR TO BE THE RESULT OF A PERSONNEL, PROCESS OR MATERIAL ISSUE. SINCE A SAMPLE WAS NOT PROVIDED, THE NON-CONFORMITY COULD NOT BE CONFIRMED.
Description of Event or Problem
SURGEON STATES THERE IS A LEAK IN SOME OF THE DRAPES. HE HAS FOUND A SMALL AMOUNT OF FLUID (LESS THAN 50CC) BETWEEN THE DRAPE AND FLUID WARMER IN TWO CASES. HE IS CONCERNED THERE IS A BREAK IN STERILITY.