Received Jun 14, 2016 · Event occurred May 10, 2016
Report 9611174-2016-00002 · MDR key 5722919
Device
Generic name
Wm-np1 Mobile Workstation
Manufacturer
Keymed (medical & Industrial Equipment) LtdModel number
WM-NP1Catalog number
K1000287
Product problems
- Device Operates Differently Than Expected
- Device Operates Differently Than Expected
Patient
Not reported
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
DEVICE WAS MANUFACTURED IN 2006. OLYMPUS (B)(4) PLAN TO RETURN THE SUBJECT DEVICE TO THE MANUFACTURER - KEYMED (MEDICAL & INDUSTRIAL EQUIPMENT) LTD FOR INVESTIGATION. KEYMED HAVE REQUESTED A COPY OF THE MAINTENANCE RECORDS FROM THE HEALTHCARE FACILITY AND ANY SERVICE EVENTS AT THE OLYMPUS REPAIR CENTRE. DEVICE AWAITING RETURN OF DEVICE.
Description of Event or Problem
THE OLYMPUS WM-NP1 SERIES MOBILE WORKSTATION IS INTENDED FOR USE IN MEDICAL FACILITIES UNDER THE DIRECTION OF A TRAINED PHYSICIAN AND HAS BEEN DESIGNED TO BE USED WITH A RANGE OF OLYMPUS EQUIPMENT TO FACILITATE GI ENDOSCOPY, ENDOSCOPY ULTRASOUND, RESPIRATORY AND SURGICAL ENDOSCOPIC PROCEDURES. KEYMED (MEDICAL & INDUSTRIAL EQUIPMENT) LTD HAS BEEN MADE AWARE OF AN EVENT WHICH TOOK PLACE IN (B)(6). THE THREADED STUD OF THE RIGHT REAR CASTOR OF THE WORKSTATION BROKE WHEN THE WORKSTATION WAS MANOEUVRED. THE WORKSTATION WAS CAUGHT BY HEALTHCARE FACILITY STAFF. THERE IS NO REPORT OF INJURY TO PATIENT OR USERS AND THIS REPORT IS SUBMITTED IN AN ABUNDANCE OF CAUTION.
Additional Manufacturer Narrative
INVESTIGATION BASED ON THE RETURN OF THE WORKSTATION CASTOR TO THE MANUFACTURER. THE WORKSTATION WAS MANUFACTURED IN 2006. THE MOST PROBABLE CAUSE IDENTIFIED BY THE INVESTIGATING ENGINEER WAS THAT THE CASTOR WASHER INSIDE DIAMETER WAS NOT SUFFICIENT TO ACCOMODATE THE CASTOR STEM. THE INSIDE DIAMETER OF THE WASHER WAS INCREASED BY 0.3MM FOLLOWING A DESIGN REVIEW IN 2007