inforMED
MalfunctionEYN

NV LARGE BORE Y-IRRIGATION SET

Received Jan 25, 2005 · Event occurred Jan 4, 2005

Report 9615050-2005-00014 · MDR key 571155

Device

Generic name

Irrigation Set

Model number

NA

Catalog number

06599

Lot number

201295H

Product problems

  • Tip breakage

Patient

67 YR

  • Surgical procedure
  • Foreign body, removal of

Narrative

Description of Event or Problem

PATIENT WAS UNDERGOING A TRANSURETHRAL RESECTION OF THE PROSTATE. THE CUSTOMER REPORTED THAT 3L GLYCINE BAGS WERE BEING CONSECUTIVELY SPIKED AND REPLACED BY THE CIRCULATING NURSE WHILE GLYCINE WAS DELIVERED VIA GRAVITY. THEIR ROUTINE IS TO CHANGE BAGS, OPEN THE FLOW WIDE OPEN, CLAMP OFF THE EMPTY LINE, AND SPIKE THE NEXT BAG. AFTER CHANGING BAG NUMBER 0 OR 10, THE SURGEON VISUALIZED A FOREIGN FRAGMENT, AND WAS ABLE TO REMOVE IT FROM THE BLADDER WITHOUT SIGNIFICANT DELAY. SUBSEQUENTLY, THE FRAGMENT WAS MATCHED UP WITH A BROKEN SECTION IN THE PIERCING PIN TIP AREA. THERE WAS NO ADVERSE PATIENT EFFECT.

Description of Event or Problem

PATIENT WAS UNDERGOING A TRANSURETHRAL RESECTION OF THE PROSTATE. THE CUSTOMER REPORTED THAT 3L GLYCINE BAGS WERE BEING CONSECUTIVELY SPIKED AND REPLACED BY THE CIRCULATING NURSE WHILE GLYCINE WAS DELIVERED VIA GRAVITY. THEIR ROUTINE IS TO CHANGE BAGS, OPEN THE FLOW WIDE OPEN, CLAMP OFF THE EMPTY LINE, AND SPIKE THE NEXT BAG. AFTER CHANGING BAG NUMBER 9 OR 10, THE SURGEON VISUALIZED A FOREIGN FRAGMENT, AND WAS ABLE TO REMOVE IT FROM THE BLADDER WITHOUT SIGNIFICANT DELAY. SUBSEQUENTLY, THE FRAGMENT WAS MATCHED UP WITH A BROKEN SECTION IN THE PIERCING PIN TIP AREA. THERE WAS NO ADVERSE PATIENT EFFECT.