inforMED
MalfunctionOIU

ARCHITECT HE4

Received Jun 9, 2016

Report 3002809144-2016-00035 · MDR key 5710065

Device

Generic name

He4

Manufacturer

Abbott Germany

Catalog number

02P54-25

Lot number

50440M500

Product problems

  • Low Test Results

Patient

40 YR

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

FURTHER INVESTIGATION OF THE CUSTOMER ISSUE INCLUDED A REVIEW OF THE COMPLAINT TEXT, A SEARCH FOR SIMILAR COMPLAINTS, REVIEW OF INSTRUMENT LOGS, A REVIEW OF LABELING AND ACCURACY TESTING. NO ADVERSE TREND WAS IDENTIFIED FOR THE CUSTOMER'S ISSUE. LABELING WAS REVIEWED AND FOUND TO BE ADEQUATE. THE PRODUCT WAS NOT AVAILABLE FOR RETURN. AN INTERNAL PANEL WAS TESTED WITH RETAINED KITS OF THE LIKELY CAUSE REAGENT LOT AND ACCURACY TESTING MET ALL SPECIFICATIONS. BASED ON ALL AVAILABLE INFORMATION AND ABBOTT DIAGNOSTICS' COMPLAINT INVESTIGATION, THE ASSAY PERFORMED AS INTENDED AND NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Additional Manufacturer Narrative

AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. (B)(4).

Description of Event or Problem

THE CUSTOMER OBSERVED A FALSE NEGATIVE HE4 RESULT ON THE ARCHITECT I1000SR ANALYZER. THE FOLLOWING DATA WAS PROVIDED: INITIAL 47.5, REPEAT 67.3 PMOL/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Additional Manufacturer Narrative

THE CONCLUSION CODE WAS CORRECTED TO (B)(4). THE DEVICE EVALUATION WAS REASSESSED AND CONCLUDED THAT A MALFUNCTION OCCURRED, THEREFORE, THE DEVICE WAS NOT PERFORMING AS INTENDED, HOWEVER, NO SYSTEMIC ISSUE OR PRODUCT DEFICIENCY WAS IDENTIFIED.