ORS-300N SOLUTION WARMER DRAPE

Received May 31, 2016 · Event occurred Apr 20, 2016

Report 8043817-2016-00033 · MDR key 5688833

Device

Generic name

Fluid Warming And Slush Drapes

Model number

ORS-300N

Lot number

SEE H10

Product problems

  • Hole In Material

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THIS EVENT WAS ASSOCIATED WITH ITEM NUMBER OM-ORS-300 LOT NUMBER 1057020 AND IS ASSOCIATED WITH THE POTENTIAL FOLLOWING ORS-300N NON STERILE MICROTEK LOT NUMBERS, DA153151, D153151, D153151A AND D153061A. AFTER EXAMINING THE RETURNED DEVICE, THE "HOLE" IN QUESTION APPEARS TO BE A SMALL BURN HOLE. BASED ON THE DEVICE HISTORY RECORD AND SAMPLE REVIEW, THE NON CONFORMITY DOES NOT APPEAR TO BE THE RESULT OF A MANUFACTURING OR MATERIAL ISSUE. MELTS, THOUGH UNCOMMON, CAN OCCUR WHEN THE WARMER IS NOT DRAPED PROPERLY, NOT TURNED OFF BY USING THE POWER BUTTON, AND/OR THERE IS INSUFFICIENT FLUID IN THE WARMER BASIN, CONTRARY TO THE OPERATIONS MANUAL, PRODUCT LABELING, PRODUCT INSERT AND IN-SERVICE PRESENTATIONS. BASED UPON THE EXAMINATION OF THE RETURNED PRODUCT, IT IS CONCLUDED THAT THIS EVENT WAS RELATED TO MISUSE BY THE CUSTOMER AND THEREFORE NO ADDITIONAL ACTIONS WILL BE TAKEN AT THIS TIME.

Description of Event or Problem

AFTER THE SURGICAL CASE WAS FINISHED, STAFF NOTICED THERE WAS SALINE BETWEEN THE DRAPE AND WARMER BASIN. UPON FURTHER INSPECTION THEY DETECTED A SMALL HOLE IN THE DRAPE. BIOMED INSPECTED THE WARMING UNIT AND FOUND IT TO PERFORM CORRECTLY.