I-STAT B-HCG CARTRIDGE
Received May 20, 2016 · Event occurred May 10, 2016
Report 2245578-2016-00032 · MDR key 5669713
Device
Generic name
B-hcg Cartridge
Manufacturer
Abbott Point Of CareModel number
NACatalog number
05P58-25
Lot number
A16044
Product problems
- False Positive Result
Patient
27 YR
- Abdominal Pain
- Abdominal Pain
Narrative
Additional Manufacturer Narrative
APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 06/02/2016. RETAIN PRODUCT WAS TESTED AND IS FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.
Description of Event or Problem
NA
Additional Manufacturer Narrative
(B)(4)
Description of Event or Problem
ON (B)(6) 2016, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A SUSPECTED FALSE POSITIVE RESULT ON A (B)(6) PATIENT WITH ABDOMINAL PAIN. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. THE PATIENT WAS GIVEN AN ULTRASOUND, WHICH CONFIRMED THE PATIENT WAS NOT PREGNANT. DATE METHOD B-HCG DATE: (B)(6) 2016, METHOD: I-STAT, B-HCG: 42.8. ON (B)(6) 2016, VISTA, <1.0. BASED ON THE INFORMATION AVAILABLE AND INTERNAL CONTROL ALGORITHMS THE EVENT IS REPORTABLE AS A POTENTIAL PRODUCT MALFUNCTION ALTHOUGH NOT CONFIRMED AT THIS TIME. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.