Additional Manufacturer Narrative

(B)(4).

Description of Event or Problem

ON (B)(6) 2016, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT B-HCG CARTRIDGES THAT YIELDED A SUSPECTED FALSE POSITIVE RESULT ON A (B)(6) YEAR OLD FEMALE PATIENT ADMITTED WITH DIABETIC FLU LIKE SYMPTOMS. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. THE PATIENT WAS GIVEN A TRANSVAGINAL ULTRASOUND, WHICH CONFIRMED THE PATIENT WAS NOT PREGNANT. THE CUSTOMER STATES THAT RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. DATE: (B)(6) 2016, TIME: 21:12, I-STAT : 44, LAB: <1. BASED ON THE INFORMATION AVAILABLE AND INTERNAL CONTROL ALGORITHMS THE EVENT IS REPORTABLE AS A POTENTIAL PRODUCT MALFUNCTION ALTHOUGH NOT CONFIRMED AT THIS TIME. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. THE INVESTIGATION IS UNDERWAY.

Additional Manufacturer Narrative

APOC INCIDENT # (B)(4). THE INVESTIGATION WAS COMPLETED ON 06/22/2016. RETAIN PRODUCT WAS TESTED AND IS FUNCTIONING ACCORDING TO SPECIFICATION. RETURN PRODUCT WAS NOT AVAILABLE FOR INVESTIGATION.

Description of Event or Problem

NA