XLTEK BRAIN MONITOR
Received May 13, 2016 · Event occurred Mar 4, 2016
Report 5652286 · MDR key 5652286
Device
Generic name
Amplitude-integrated Electroencephalograph
Manufacturer
NatusProduct problems
- Device Alarm System
- Computer Software Problem
Patient
Not reported
- Seizures
- Seizures
Narrative
Description of Event or Problem
PATIENT WAS BEING MONITORED ON EEG FOR POTENTIAL SEIZURE ACTIVITY. THE ALARM FROM THE EEG SYSTEM DID NOT ALERT AS EXPECTED BY CLINICIANS. INVESTIGATIVE FINDINGS INDICATED THAT THE SOFTWARE THAT ELICITS THE ALARM WENT INTO A PAUSE STATE TEMPORARILY. THE SOFTWARE HAS SINCE BEEN UPDATED AND ACCORDING TO PERSYST THE PATCH SHOULD PREVENT A RE-OCCURRENCE. MANUFACTURER RESPONSE FOR EEG ACQUISITION MODULE, NA (PER SITE REPORTER): THE MANUFACTURER OF THE SOFTWARE PERSYST ANALYZED ERROR LOGS AND FINDINGS INDICATED A PROBLEM THAT THEY COULD NOT REPLICATE IN THEIR ENVIRONMENT. HOWEVER, THEY HAVE PROVIDED A PATCH TO ADDRESS THE PROBLEM