inforMED
OtherFAM

B-K MEDICAL

Received Dec 2, 2004 · Event occurred Nov 22, 2004

Report 559053 · MDR key 559053

Device

Generic name

Rectosigmoidoscope

Model number

DO 0163

Catalog number

*

Lot number

*

Product problems

  • Difficult to Remove
  • Increase in Suction
  • Physical Resistance

Patient

*

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

DOCTOR WAS USING THE RECTOSIGMOIDOSCOPE TO PROVIDE ACCESS FOR A TYPE 2050 ANORECTAL ULTRASOUND TRANSDUCER. DOCTOR NOTICED THAT WHEN ATTEMPTING TO REMOVE THE OBTURATOR, THERE WAS SIGNIFICANT RESISTANCE, INDICATING THE OBTURATOR WAS PULLING A VACUUM. DOCTOR FEELS THERE IS A RISK OF PULLING PART OF THE INTESTINE INTO THE RECTOSIGMOIDOSCOPE AND INJURING THE PATIENT.IT IS OUR OPINION THAT SOME KIND OF CHANNEL OR VENT IS NEEDED TO PREVENT A VACUUM WHEN THE OBTURATOR IS REMOVED. THERE WAS NO PATIENT INJURY.