inforMED
MalfunctionKEY

VP 2000 PROCESSOR

Received Apr 15, 2016 · Event occurred Mar 17, 2016

Report 3005248192-2016-00002 · MDR key 5583428

Device

Generic name

Automated Slide Stainer

Catalog number

02J11-01

Product problems

  • Temperature Problem
  • Overheating of Device
  • Temperature Problem

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

AN MDR FOLLOW-UP REPORT WILL BE SUBMITTED AFTER THE ELEVATED COMPLAINT INVESTIGATION CONCLUDES.

Description of Event or Problem

THE ABBOTT VP 2000 PROCESSOR IS A DEVICE DESIGNED TO AUTOMATE AND STANDARDIZE SLIDE SPECIMEN PROCESSING AND ROUTINE SLIDE STAINING FOR THE LABORATORY. CUSTOMER REPORTED THAT WHEN SLIDES WERE REMOVED FROM THE DRYING STATION ON THE VP2000 INSTRUMENT THE SLIDE RACK WAS NOTICEABLY HOTTER THAN USUAL. THERE WAS NO REPORT OF INJURY. THE ABBOTT FIELD SERVICE ENGINEER (FSE) CLEANED OFF SIGNS OF DRIP/DRY BUFFER ON AND AROUND THE DRYER ASSEMBLY THERMISTOR. TEMPERATURE VERIFICATIONS AND CHECKS WERE WITHIN MANUFACTURER SPECIFICATIONS. THE FSE RESTORED THE VP2000 TO ITS OPERATIONAL CONDITION. AN ABBOTT ELEVATED COMPLAINT INVESTIGATION HAS BEEN INITIATED.

Additional Manufacturer Narrative

IT WAS IDENTIFIED ON APRIL 20, 2017, THAT UPON SUBMISSION OF THE INITIAL REPORT, DATE OF THIS REPORT INCORRECTLY STATED 03/17/2016 WHEN IT SHOULD HAVE STATED 04/15/2016.

Additional Manufacturer Narrative

ELEVATED COMPLAINT INVESTIGATION (B)(4) FOR MDR 3005248192-2016-00002 FOLLOW-UP REPORT 1 INCLUDED AN EXISTING DATA REVIEW, QUALITY DATA REVIEW (CAPA/NONCONFORMANCE REVIEW), PRODUCT/SYSTEM/INSTRUMENT EVALUATION, AND COMPLAINT HISTORY REVIEW. THE INFORMATION IN THIS COMPLAINT IDENTIFIED THAT DRYER ASSEMBLY (PART 30-144163RO) FAILURES DID OCCUR HOWEVER THE OVERALL PRODUCT DESIGN REQUIREMENTS CONTINUE TO BE MET. THE OVERALL PRODUCT PERFORMED AS DESIGNED, PER SAFETY STANDARD IEC 61010-1, AND CONTINUES TO PERFORM AS EXPECTED WHEN THIS FAILURE MODE OCCURS, AS THERE IS NO EVIDENCE THAT THE SINGLE FAULT CONDITION TEMPERATURE EXCEEDED 105°C ON ANY TOUCHABLE SURFACE. BASED ON THE INVESTIGATION RESULTS, NO PRODUCT DEFICIENCY WAS IDENTIFIED.