inforMED
OtherMOS

PHASED ARRAY WRIST COIL

Received Oct 27, 2004 · Event occurred Oct 26, 2004

Report 556935 · MDR key 556935

Device

Generic name

Mri Extremity Coil

Model number

455GE-64

Catalog number

OPERATOR'S MANUAL 5/29/02 REV

Lot number

NOT KNOWN

Product problems

  • Unknown (for use when the device problem is not known)

Patient

69 YR

  • Burn(s)

Narrative

Description of Event or Problem

PATIENT RECEIVED 1ST DEGREE BURNS (MILD SUNBURN-LIKE APPEARANCE) TO THE RIGHT WRIST, RIGHT HUMERUS AREA (UNDERARM) AND THE 2ND AND 3RD DIGITS OF THE RIGHT HAND DURING AN MRI STUDY. PATIENT WAS SWEATING DURING TESTING. THEY WERE ALSO MAKING CONTACT WITH THE MACHINE DURING SOME OF THE EXAMINATION. GE ENGINEER EXAMINED THE WRIST COILS OF THE MACHINE AND CHECKED CALIBRATION. NO PROBLEMS WERE FOUND APPLICABLE TO THE INJURY. SOME METAL SHAVINGS WERE FOUND IN THE COIL, BUT THIS WAS DETERMINED TO NOT BE APPLICABLE TO THE INJURY.