inforMED
MalfunctionKRR

GENDER SOLUTIONS PATELLO-FEMORAL TROCHLEA

Received Apr 11, 2016 · Event occurred Jun 22, 2015

Report 1822565-2016-01099 · MDR key 5567247

Device

Generic name

Krr

Manufacturer

Zimmer Inc

Catalog number

00592601102

Lot number

62054510

Product problems

  • Nonstandard Device
  • No Pressure

Patient

67 YR

  • No Information
  • No Information

Narrative

Additional Manufacturer Narrative

(B)(4). NO DEVICES OR PHOTOS WERE RECEIVED; THEREFORE THE CONDITION OF THE COMPONENTS IS UNKNOWN. THIS DEVICE IS USED FOR TREATMENT. THIS ISSUE HAS BEEN PREVIOUSLY INVESTIGATED AND AS PART OF CORRECTIVE ACTION, A NEW SUPPLIER FOR THE POLYETHYLENE BAGS HAS BEEN SELECTED AND TESTING COMPLETED TO ENSURE THERMAL RELIABILITY AND STABILITY OF THE NEW BAGS. FIELD ACTION Z-0845-2016 WAS INITIATED ON JANUARY 11, 2016 TO REMOVE REMAINING AFFECTED UNITS FROM THE FIELD. THIS FIELD ACTION CONTAINS THE RELATED PART AND LOT NUMBER. THIS MDR WAS IDENTIFIED DURING AN INTERNAL RETROSPECTIVE REVIEW TO MEET REPORTING REQUIREMENTS AND THEREFORE IS BEING FILED RETROSPECTIVELY.

Description of Event or Problem

IT IS REPORTED THAT THE INNER PLASTIC PACKAGING ADHERED TO THE SURFACE OF THE IMPLANT.

Remedial action

  • Recall