inforMED
MalfunctionFPL

AIRBORNE TRANSPORT INCUBATORS A 750I

Received Apr 6, 2016 · Event occurred Mar 25, 2016

Report 5549836 · MDR key 5549836

Device

Generic name

Incubator, Neonatal Transport

Model number

A 750I

Product problems

  • Thermal Decomposition of Device

Patient

Not reported

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Description of Event or Problem

THE ORANGE WIRE RUNNING FROM THE BRIDGE RECTIFIER TO THE TRANSFORMER BURNT UP. ALL THE WIRES AND INTERNAL COMPONENTS ARE ORIGINAL FROM THE FACTORY. THIS CAUSED A POWER FAILURE AND THE ISOLETTE WAS UNABLE TO FUNCTION.