GALILEO INTRAVASCULAR RADIOTHERAPY CARTRIDGE, SOURCE WIRE

Received Apr 21, 2004 · Event occurred Dec 2, 2003

Report 1651662-2004-00004 · MDR key 554040

Device

Generic name

Radiation System Cartridge

Manufacturer

Guidant, Houston

Model number

1008112-0C

Catalog number

NA

Lot number

NA

Product problems

  • Failure to Power Up

Patient

UNKNOWN

  • Radiation Underdose

Narrative

Description of Event or Problem

THE PHYSICIST REPORTED THE FOLLOWING SCENARIO DURING A TREATMENT: THE SITE STARTED TO TREAT THE FIRST POSITION IN A TWO-POSITION TREATMENT. THE SYSTEM LOST POWER DURING THIS TREATMENT; THEREFORE, THE FIRST POSITION ONLY RECEIVED 1973 CGY DOSE. INSTEAD OF CONTINUING THIS CASE, THE ONCOLOGIST REPOSITIONED THE CATHETER TO THE SECOND POSITION AND STARTED A NEW TWO-POSITION TREATMENT. THE SECOND POSITION RECEIVED THE NORMAL 20 GY DOSE AND THEN THE SYSTEM RETRACTED TO THE THIRD POSITION AND DWELLED IN THIS THIRD POSITION FOR A COUPLE OF SECONDS BEFORE THE TREATMENT. THE SECOND POSITION RECEIVED THE NORMAL 20 GY DOSE AND THEN THE SYSTEM RETRACTED TO THE THIRD POSITION AND DWELLED IN THIS THIRD POSITION FOR A COUPLE OF SECONDS BEFORE THE TREATMENT WAS INTERRPUTED. THIS THIRD AREA RECEIVED A 56 CGY DOSE.