inforMED
InjuryFYF

BOUFFANT CAP

Received Mar 29, 2016

Report 1417592-2016-00025 · MDR key 5532195

Device

Generic name

Bouffant Cap

Catalog number

CRI1004Z

Lot number

26315060024

Product problems

  • Patient-Device Incompatibility

Patient

25 YR

  • Irritation
  • Irritation

Narrative

Additional Manufacturer Narrative

IT WAS REPORTED THAT THREE STUDENTS DEVELOPED EYE IRRITATION AND SKIN IRRITATION ON THEIR FOREHEAD AFTER WEARING THIS CAP. ONE STUDENT WAS TREATED WITH AN ANTIBIOTIC OINTMENT. THE OTHER TWO STUDENTS WERE TREATED WITH AN OTC EYE DROP. THE ETIOLOGY OF THE EYE IRRITATION IS UNKNOWN. UNUSED SAMPLES FROM THE SAME LOT WERE RETURNED AND SENT TO AN INDEPENDENT LAB. THE TEST RESULTS REVEALED THAT THE SAMPLES WERE NON-CYTOTOXIC. A ROOT CAUSE HAS NOT BEEN DETERMINED. IT IS UNKNOWN IF THE BOUFFANT CAP CAUSED OR CONTRIBUTED TO THE REPORTED SKIN AND EYE IRRITATION. HOWEVER, DUE TO THE NEED FOR MEDICAL INTERVENTION AND IN AN ABUNDANCE OF CAUTION, THIS MEDWATCH IS BEING FILED.

Description of Event or Problem

IT WAS REPORTED THAT THREE STUDENTS DEVELOPED EYE IRRITATION AND SKIN IRRITATION ON THEIR FOREHEAD AFTER WEARING THIS CAP.