FMP HIP
Received Mar 29, 2016 · Event occurred Mar 4, 2016
Report 1644408-2016-00219 · MDR key 5531713
Device
Generic name
Liner/non-hooded-neu, Mp10, Hxe-plus, 44mm
Manufacturer
Encore Medical, L.p.Catalog number
931-44-760
Lot number
764N1002
Product problems
- Device Dislodged or Dislocated
Patient
74 YR
- Joint Dislocation
- Joint Dislocation
Narrative
Additional Manufacturer Narrative
THE REASON FOR THIS REVISION SURGERY WAS HIP DISLOCATION. THE LENGTH OF THE IN-VIVO SERVICE WAS 2.4 MONTHS. THERE IS NO INFORMATION IN THIS COMPLAINT ABOUT ANY PATIENT INJURIES, ACTIVITIES, OR ACCIDENTS THAT MAY HAVE CONTRIBUTED TO THE NEED FOR THIS REVISION SURGERY. THE HEALTHCARE PROFESSIONAL INDICATED THERE WAS A SERIOUS RISK TO THE PATIENT. THERE WAS NO DELAY IN SURGERY AND ANOTHER SUITABLE DEVICE WAS AVAILABLE FOR USE. THE REVISION SURGERY WAS COMPLETED AS INTENDED. THE DEVICE WAS DISPOSED OF AT THE HOSPITAL AND NOT MADE AVAILABLE TO (B)(4) FOR EXAMINATION. A REVIEW OF THE DEVICE HISTORY RECORDS (DHRS) REVEALED NO DISCREPANCIES OR ISSUES WITH THE MANUFACTURING HISTORY OF THIS PART. ALL PARTS WERE FOUND TO MEET DESIGN AND MANUFACTURING SPECIFICATIONS. NO NON-CONFORMING MATERIAL REPORTS (NCMR'S) WERE ASSOCIATED WITH THIS PART. THE PRODUCT COMPLAINT REPORT HISTORY WAS REVIEWED AND NO TRENDS OR ON-GOING ISSUES WERE DEEMED AS PRESENT OR IN NEED OF REVIEW. THIS EVENT IS DEEMED TO BE NON-PRODUCT RELATED. THE SURGEON REPORTED NO ISSUES ASSOCIATED WITH THE EXPLANTED PRODUCT AND PROVIDED NO INFORMATION THAT DEFINITIVELY DESCRIBED THE ROOT CAUSE OR REASON OF THE DISLOCATION. THE SCOPE OF THIS INVESTIGATION IS LIMITED WITHOUT HAVING THE PARTS AVAILABLE TO (B)(4)FOR EVALUATION. OTHER CONDITIONS RELATING TO THIS EVENT COULD NOT BE DETERMINED WITH CONFIDENCE. FACTORS THAT MAY CONTRIBUTE TO A DISLOCATION NOT ASSOCIATED WITH THE IMPLANT ARE: IMPROPER IMPLANT SELECTION, DEGENERATIVE BONE DISEASE, AND IMPROPER SURGICAL TECHNIQUE. INVENTORY CONTAINMENT IS NOT REQUIRED AS THERE ARE NO INDICATIONS OF A PRODUCT OR PROCESS ISSUE AFFECTING IMPLANT SAFETY OR EFFECTIVENESS.
Description of Event or Problem
REVISION SURGERY - DUE TO THE PATIENT DISLOCATING.