MalfunctionKSO
ID-MTS READER SA
Received Apr 19, 2004 · Event occurred Apr 2, 2004
Report 1058138-2004-00014 · MDR key 552354
Device
Generic name
Blood Grouping Centrifuge And Reader
Manufacturer
Micro Typing Systems, Inc.Model number
MTS9945Catalog number
MTS READER SA
Lot number
*
Product problems
- Break
Patient
NA
- Unknown (for use when the patient's condition is not known)
Narrative
Description of Event or Problem
LID ON READER SA COULD SLAM SHUT IF HINGE ASSEMBLY IS BROKEN. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT.
Remedial action
- Other
- Repair