inforMED
MalfunctionIOY

SUPPORT ARM 177

Received Mar 18, 2016

Report 8010042-2016-00152 · MDR key 5510245

Device

Generic name

Support, Arm

Catalog number

6481720

Product problems

  • Break

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Additional Manufacturer Narrative

OUR INVESTIGATION INTO THIS COMPLAINT HAS NOW BEEN FINALIZED. NO PARTS OR PICTURES WERE RETURNED FOR OUR INVESTIGATION. ADDITIONAL INFORMATION RECEIVED CLARIFIED THAT THE SUPPORT ARM WILL NOT REMAIN IN ITS INTENDED/DESIRED POSITION. OUR CONCLUSION IS, THAT THE FRICTION IN THE JOINT HAS BEEN REDUCED, MOST LIKELY DUE TO WEAR. THERE WAS/IS NO EVIDENCE TO SUPPORT A MECHANICAL BREAK OF THE DEVICE. THERE ARE SERVICE KITS AVAILABLE FOR EXCHANGING PARTS OF THE JOINTS WHEN THEY BECOME WORN OUT.

Description of Event or Problem

(B)(4).

Description of Event or Problem

(B)(6) 2016 09:00 AM (GMT-4:00) ADDED BY (B)(6) ((B)(4)): IT WAS REPORTED THAT THE SUPPORT ARM FOR THE VENTILATOR BROKE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. (B)(4).