SIMPLEXA HSV 1 & 2 DIRECT, SIMPLEXA HSV 1 & 2 POSITIVE CONTROL PACK
Received Mar 10, 2016 · Event occurred Jan 11, 2016
Report 2023365-2016-00031 · MDR key 5493882
Device
Generic name
Herpes Simplex Virus Nucleic Acid Amplification Assay
Manufacturer
Focus DiagnosticsModel number
MOL2150Catalog number
NA
Lot number
29813
Product problems
- Incorrect Or Inadequate Test Results
- Incorrect Or Inadequate Test Results
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
FOCUS DIAGNOSTICS HAS RECEIVED A TOTAL OF 34 COMPLAINTS FROM 20 CUSTOMERS FOR EARLY CYCLE THRESHOLD (CT) ISSUES WITH THE DIRECT AMPLIFICATION DISC (DAD) (MOL1455, MOL1451, AND MOL 1452) WHILE PERFORMING THE (B)(4) DIRECT ASSAY. THIS ERROR MAY RESULT IN A (B)(6) RESULT, AND ERROR CODE(S) GIVING AN INVALID RUN. EARLY INTERNAL TESTING REPRODUCED THE EARLY CT THRESHOLD; HOWEVER, THESE EARLY CT WERE PRODUCED ONLY WHEN THE DAD WAS USED AFTER A PREVIOUS PARTIAL RUN. THE FALSE RESULT WAS NOT REPORTED TO THE DOCTOR. THERE WAS NO PATIENT INVOLVEMENT. A VOLUNTARY RECALL HAS BEEN PERFORMED RELATED TO THE REPORTED LOT. AS OF (B)(6) 2016, THERE HAS BEEN NO REPORT OF PATIENT INJURY OR DEATH.
Remedial action
- Notification