inforMED
MalfunctionBYT

NEV100

Received Apr 8, 2004 · Event occurred Mar 9, 2004

Report 2518422-2004-00027 · MDR key 548635

Device

Generic name

Negative Pressure External Ventilator

Manufacturer

Respironics, Inc.

Model number

12001

Catalog number

12001

Lot number

NA

Product problems

  • No Audible Alarm
  • Loose
  • Repair

Patient

NA

Narrative

Description of Event or Problem

THE UNIT WAS RETURNED FOR PREVENTATIVE MAINTENANCE WITH NO PROBLEMS IDENTIFIED. THE UNIT HAD NOT BEEN SERVICED BY RESPIRONICS SINCE 2000. DURING THE REPAIR EVAL IT WAS DISCOVERED THAT THE UNIT'S AUDIBLE ALARM WAS NOT FUNCTIONING. THE DISPLAY AND POWER BOARDS HAD BECOME LOOSE AND WERE RESEATED TO CORRECT THE PROBLEM. THERE WAS NO PT INVOLVEMENT OR REPORTED HARM.

Remedial action

  • Other