SIMPLEXA HSV 1 & 2 DIRECT, SIMPLEXA HSV 1 & 2 POSITIVE CONTROL PACK
Received Mar 6, 2016 · Event occurred Jan 6, 2016
Report 2023365-2016-00011 · MDR key 5482717
Device
Generic name
Herpes Simplex Virus Nucleic Acid Amplification Assay
Manufacturer
Focus DiagnosticsModel number
MOL2150Catalog number
NA
Lot number
29691
Product problems
- False Positive Result
- False Positive Result
Patient
Not reported
- No Consequences Or Impact To Patient
- No Consequences Or Impact To Patient
Narrative
Description of Event or Problem
FOCUS DIAGNOSTICS HAS RECEIVED A TOTAL OF 34 COMPLAINTS FROM 20 CUSTOMERS FOR EARLY CYCLE THRESHOLD (CT) ISSUES WITH THE DIRECT AMPLIFICATION DISC (DAD) (MOL1455, MOL1451, AND MOL 1452) WHILE PERFORMING THE SIMPLEXA HSV 1 & 2 DIRECT ASSAY. THIS ERROR MAY RESULT IN A (B)(6), AND ERROR CODE(S) GIVING AN INVALID RUN. EARLY INTERNAL TESTING REPRODUCED THE EARLY CT THRESHOLD; HOWEVER, THESE EARLY CT WERE PRODUCED ONLY WHEN THE DAD WAS USED AFTER A PREVIOUS PARTIAL RUN. THIS MDR IS BEING SUBMITTED BECAUSE THE (B)(6) RESULTS WERE REPORTED TO THE DOCTOR.A VOLUNTARY RECALL HAS BEEN PERFORMED RELATED TO THE REPORTED LOT.AS OF MARCH 4, 2016, THERE HAS BEEN NO REPORT OF PATIENT INJURY OR DEATH.
Remedial action
- Notification