SUPPORT ARM 177

Received Mar 3, 2016

Report 8010042-2016-00114 · MDR key 5476244

Device

Generic name

Support, Arm

Catalog number

6481720

Product problems

  • Break

Patient

Not reported

  • No Patient Involvement
  • No Patient Involvement

Narrative

Description of Event or Problem

ON (B)(6) 2016 11:45 AM (GMT-5:00) ADDED BY (B)(6) ((B)(4)): IT WAS REPORTED THAT THE SUPPORT ARM WAS FOUND BROKEN. THERE WAS NO PATIENT INVOLVEMENT REPORTED. (B)(4).

Additional Manufacturer Narrative

THE INVESTIGATION OF THE RETURNED SUPPORT ARM HAS BEEN FINALIZED. IT WAS REPORTED THAT THE SUPPORT ARM WAS FOUND BROKEN BY OUR SERVICE ENGINEER ON ANOTHER ASSIGNMENT DURING A SERVICE VISIT AT SITE. VISUAL INSPECTION SHOWS THAT THERE IS A CLEAR RUPTURE AND IN THE MICROSCOPIC INVESTIGATION THE CONCLUSION IS THAT THERE IS NO CASTING DEFECT. THIS IMPLIES THAT THE SUPPORT ARM HAS BEEN EXPOSED TO A MECHANICAL FORCE THAT'S ABOVE THE DESIGNED SPECIFICATION. THE SUPPORT ARM HAS BEEN SUCCESSFULLY TESTED FOR MECHANICAL STRENGTH AND IS DESIGN ACCORDING TO STANDARD. IT¿S UNKNOWN TO US UNDER WHICH CONDITIONS THE REPORTED SUPPORT ARM BROKE.