WARMER DRAPE
Received Jan 25, 2016 · Event occurred Dec 14, 2015
Report 8043817-2016-00004 · MDR key 5389230
Device
Generic name
Fluid Warming And Slush Drapes
Manufacturer
Microtek Dominicana, S.a.Model number
ORS-100NLot number
UNKNOWN
Product problems
- Fluid/Blood Leak
- Hole In Material
Patient
70 YR
- No Known Impact Or Consequence To Patient
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
SAMPLE RECEIVED ON 1/11/2016, MANUFACTURER REVIEWED IT ON 1/13/2016. SAMPLE DRAPE WAS SOAKED IN BLEACH WATER FOR ABOUT AN EXTENDED PERIOD OF TIME (1.5 HR +), RINSED, AND LET DRY. CHECKED SAMPLE AND TOOK PHOTOS OF BURN MARKS AND BURN HOLE. REVIEW OF THE SAMPLE DRAPE FOUND TWO 2 INCH BURN MARKS 2 INCHES FROM THE CENTER OF THE DRAPE AND A SMALL BURN HOLE 3 INCHES FROM THE CENTER OF THE DRAPE AND 5 INCHES DOWN. A LOT NUMBER WAS NOT PROVIDED AND THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED BUT AFTER INSPECTION OF THE RETURN SAMPLE THE PRODUCT APPEARS TO BE MANUFACTURED TO SPECIFICATION. THIS NON CONFORMITY DOES NOT APPEAR TO BE THE RESULT OF A MATERIAL ISSUE. AFTER REVIEWING THE SAMPLE, IT IS CONCLUDED THAT THE HOLES APPEAR TO BE MELTS OR BURNS IN THE DRAPE. THE BURN MARKS AND BURN HOLE DO NOT APPEAR TO BE THE RESULT OF A PERSONNEL, PROCESS, OR MATERIAL ERROR. MELTS CAN OCCUR WHEN THE WARMER IS NOT DRAPED PROPERLY, NOT TURNED OFF BY USING THE POWER BUTTON, AND/OR THERE IS INSUFFICIENT FLUID IN THE WARMER BASIN, CONTRARY TO THE OPERATIONS MANUAL, PRODUCT LABELING, PRODUCT INSERT AND IN-SERVICE PRESENTATIONS.
Description of Event or Problem
AFTER THE CASE WAS DONE, STAFF NOTICED THAT THERE WAS WATER ON THE BOTTOM OF THE SOLUTIONS WARMER. THE STAFF NOTED THAT THERE WAS A HOLE IN THE ACTUAL DRAPE. THE SURGEON WAS NOTIFIED AND THE PROCEDURE STANDARD OF CARE WAS CHANGED FROM "CLEAN CONTAMINATED" TO "CONTAMINATED."
Additional Manufacturer Narrative
SAMPLE RECEIVED ON 01/11/2016, MANUFACTURER REVIEWED IT ON 01/13/2016. SAMPLE DRAPE WAS SOAKED BLEACH WATER FOR ABOUT AN EXTENDED PERIOD OF TIME (1.5 HR. +), RINSED AND LET DRY. CHECKED SAMPLE AND TOOK PHOTOS OF BURN MARKS AND BURN HOLE. REVIEW OF THE SAMPLE DRAPE FOUND TWO 2 INCH BURN MARKS 2 INCHES FROM THE CENTER OF THE DRAPE AND A SMALL BURN HOLE 3 INCHES FROM THE CENTER OF THE DRAPE AND 5 INCHES DOWN. A LOT NUMBER WAS NOT PROVIDED AND THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED BUT AFTER INSPECTION OF THE RETURN SAMPLE THE PRODUCT APPEARS TO BE MANUFACTURED TO SPECIFICATION. THIS NON CONFORMITY DOES NOT APPEAR TO BE THE RESULT OF A MATERIAL ISSUE. AFTER REVIEWING THE SAMPLE, IT IS CONCLUDED THAT THE HOLES APPEAR TO BE MELTS OR BURNS IN THE DRAPE. THE BURN MARKS AND BURN HOLE DO NOT APPEAR TO BE THE RESULT OF A PERSONNEL, PROCESS OR MATERIAL ERROR. MELTS CAN OCCUR WHEN THE WARMER IS NOT DRAPED PROPERLY, NOT TURNED OFF BY USING THE POWER BUTTON, AND/OR THERE IS INSUFFICIENT FLUID IN THE WARMER BASIN, CONTRARY TO THE OPERATIONS MANUAL, PRODUCT LABELING, PRODUCT INSERT AND IN-SERVICE PRESENTATIONS.
Description of Event or Problem
AFTER THE CASE WAS DONE, STAFF NOTICED THAT THERE WAS WATER ON THE BOTTOM OF THE SOLUTIONS WARMER. THE STAFF NOTED THAT THERE WAS A HOLE IN THE ACTUAL DRAPE. THE SURGEON WAS NOTIFIED AND THE PROCEDURE STANDARD OF CARE WAS CHANGED FROM "CLEAN CONTAMINATED" TO "CONTAMINATED."