inforMED
OtherFJD

BD PARADIGM LINK GLUCOSE MONITOR

Received Sep 10, 2004 · Event occurred Aug 12, 2004

Report 2243072-2004-00548 · MDR key 5351803

Device

Generic name

Bd Test Strips

Manufacturer

Becton Dickinson

Model number

NA

Catalog number

322201

Lot number

3064100

Product problems

  • High Readings

Patient

UNK

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.

Description of Event or Problem

CONSUMER CALLED TO REPORT GETTING HIGH READINGS W/HER PLINK METER. TESTED AGAINST HER OTHER METER. ANALYSIS RUN ON CLARK ERROR GRID USING COMPETITIVE READING (47) AS REF. POINT VS. BD READINGS (481) YIELDED DATA POINTS IN THE E RANGE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.