Received Sep 10, 2004 · Event occurred Aug 12, 2004

Report 2243072-2004-00545 · MDR key 5351795

Device

Generic name

Bd Test Strips

Manufacturer

Becton Dickinson

Model number

NA

Catalog number

322051

Lot number

3064062

Product problems

  • False Reading From Device Non-Compliance

Patient

UNK

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

AWAITING PRODUCT RETURN TO CONDUCT QUALITY INVESTIGATION.

Description of Event or Problem

CONSUMER CALLED TO REPORT HER METER IS GIVING INACCURATE READINGS, VERY ERRATIC. ANALYSIS RUN ON CLARK ERROR GRID USING MEAN AVG. READING (212 AS REF. POINT VS. BD READINGS (36, 269, 176, 270, 311) YIELDED DATA POINTS IN THE E RANGE OF THE GRID, OUTSIDE OF ACCEPTABLE LIMITS.