ORS FLUID WARMER

Received Jan 4, 2016 · Event occurred Nov 2, 2015

Report 1043582-2016-00001 · MDR key 5342361

Device

Generic name

Warmer, Irrigation Solution

Model number

ORS-2066R

Product problems

  • Fluid/Blood Leak
  • Device Operates Differently Than Expected
  • Temperature Problem

Patient

39 YR

  • No Known Impact Or Consequence To Patient
  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE CUSTOMER WAS CONTACTED FOR ADDITIONAL INFORMATION AND WANTED TO WAIT FOR THEIR BIOMED DEPARTMENT TO CHECK OUT THE UNIT BEFORE SENDING IT FOR INSPECTION AND REPLACEMENT. NO PRODUCT HAS BEEN RETURNED. SINCE NO DEVICE WAS RETURNED FOR ANALYSIS, THE ROOT CAUSE FOR THE EVENT HAS NOT BEEN DETERMINED.

Description of Event or Problem

DURING OPERATING ROOM CLEAN UP, AFTER COMPLETION OF THE SURGERY, THE WARMING DRAPE WAS REMOVED FROM THE SOLUTION WARMER. THE SURGICAL TECHNICIAN NOTED THERE WAS LIQUID UNDER THE DRAPE AND THE TEMPERATURE ON THE WARMER HAD RISEN TO 107 DEGREES WITHOUT ANYONE CHANGING THE PRESET TEMPERATURE. THE DRAPE WAS INSPECTED. NO VISIBLE HOLE WAS FOUND. THE SURGEON WAS NOTIFIED AND THE WARMER WAS TAKEN OUT OF SERVICE.