inforMED
InjuryJAQ

VAGINAL APPLICATOR SET WITH CT CYLINDER

Received Jun 10, 2004 · Event occurred Mar 11, 2004

Report 9612638-2004-00001 · MDR key 529515

Device

Generic name

Remote Controlled Radionuclide Applicator System

Model number

11-00414

Catalog number

GM11004140

Lot number

S40

Product problems

  • Device Slipped

Patient

UNKNOWN

  • Erythema

Narrative

Description of Event or Problem

DURING A BRACHYTHERAPY TREATMENT A PT RECEIVED RADIATION ERYTHEMA ON BOTH INNER THIGHS. THE PROBE IN A VAGINAL CYLINDER 3.0CM APPLICATOR SLIPPED OUT OF THE APPLICATOR FOR 1/3 FRACTIONS (3 FRACTIONS AT 7GY AT 0.5CM). THE USER SITE CONCLUDED THAT THE FABRIC OF PT'S GARMENT WAS IN BETWEEN THE CLAMPING BOLT AND TANDEM CAUSING THE SLIP.