inforMED
InjuryJPE

MDA ANTITHROMBIN III

Received May 27, 2004 · Event occurred Apr 23, 2004

Report 3002769706-2004-00006 · MDR key 527411

Device

Generic name

Class Ii, 864.7060 Antithrombin Iii Assay

Manufacturer

Biomerieux, Inc.

Model number

*

Catalog number

252574

Lot number

UNK

Product problems

  • Use of Device Problem
  • Low Test Results

Patient

UNKNOWN

  • Treatment with medication(s)

Narrative

Description of Event or Problem

CUSTOMER CALLED TO ASK ABOUT RESULTS GIVEN WITH THE MDA ANTITHROMBIN III REAGENT (MDA ATIII). THE CUSTOMER REPORTED THAT THEY HAD QUESTIONABLY LOW RESULTS. THE CUSTOMER WAS ADVISED TO RUN WITH THE CAPS OFF RATHER THAN IN CAP PIERCING MODE. THE RESULTS WERE HIGHER AND CLOSER TO EXPECTED VALUES. THE CUSTOMER FAXED IN THE DATA WHICH SHOWED A MARKED DIFFERENCE IN RESULTS BETWEEN "CAPS ON" AND "CAPS OFF." THE CUSTOMER REPORTED THAT TWO PTS WERE PLACED ON COUMADIN BASED ON THE RESULTS WITH THE CAPS ON.

Remedial action

  • Notification