130MM X 2.3MM DIAMETER,CONCEPT HEATWAVE ELECTRODE, QTY 5
Received Nov 30, 2015 · Event occurred Nov 11, 2015
Report 1017294-2015-00074 · MDR key 5254667
Device
Generic name
Electrode, Electrosurgical
Manufacturer
Conmed LinvatecCatalog number
CS-023
Lot number
643395
Product problems
- Thermal Decomposition of Device
Patient
Not reported
- No Consequences Or Impact To Patient
Narrative
Additional Manufacturer Narrative
CONMED RECEIVED THE THREE CONCEPT HEATWAVE ELECTRODES FOR EVALUATION ON (B)(6) 2015. VISUAL INSPECTION OF THE RETURNED DEVICES CONFIRMED THE SIDE OF EACH SHAFT WAS BURNT AND MELTED IN APPEARANCE AND A PORTION OF THE COATING WAS MISSING. UPON FURTHER ANALYSIS BY R&D ENGINEERING ON 11/11/2015 USING MICROSCOPIC INSPECTION, LONGITUDINAL MARKS ALONG THE LENGTH OF THE SHRINK TUBING WERE OBSERVED. AN INVESTIGATION HAS BEEN OPENED TO ADDRESS THIS ISSUE. THIS DEVICE WAS MANUFACTURED ON 17-APR-2015 IN A LOT OF (B)(4) UNITS. (B)(4)). A REVIEW OF THE DEVICE HISTORY RECORD SHOWED NO ANOMALIES OR NON-CONFORMANCES DURING THE MANUFACTURING PROCESS. THERE HAVE NOT BEEN ANY OTHER ADVERSE EVENTS REPORTED FOR THIS DEVICE. ONE OTHER COMPLAINT REPORT WAS OPENED FOR THIS DEVICE AND LOT NUMBER COMBINATION WHEN TWO (2) ADDITIONAL UNITS WERE RETURNED UNOPENED/UNUSED. THIS FAILURE MODE IS ADDRESSED IN THE FMEA AND THE SAFETY RISK HAS BEEN FOUND TO BE ACCEPTABLE. THE PRODUCT INSERT PROVIDES THE FOLLOWING PRECAUTION: IF ANY VISUAL DEFECTS ARE NOTICED IN THE INSULATION, OR THE CERAMIC/INSULATION IS DAMAGED IN ANY WAY, STOP USING THE DEVICE IMMEDIATELY AND REPLACE THE DEVICE.
Description of Event or Problem
IT WAS FIRST REPORTED BY THE CUSTOMER ON (B)(6) 2015, THAT THREE CONCEPT HEATWAVE ELECTRODES BURNED OUT DURING A SINGLE SURGICAL PROCEDURE BEFORE THE ELECTRODE COULD BE USED ON THE PATIENT. THERE WAS NO PATIENT IMPACT. ANALYSIS OF THE RETURNED ELECTRODES BY THE R&D ENGINEER ON (B)(6) 2015, FOUND THAT THERE WAS AN INSULATION FAILURE OF THE SHRINK TUBING.