MalfunctionIWE

ALIGNRT

Received Nov 19, 2015 · Event occurred Oct 21, 2015

Report 3004832819-2015-00001 · MDR key 5249571

Device

Generic name

Patient Positioning System

Manufacturer

Vision Rt Limited

Model number

ALIGNRT

Catalog number

ALIGNRT

Lot number

NA

Product problems

  • Device Displays Incorrect Message
  • Positioning Problem

Patient

NK

  • No Known Impact Or Consequence To Patient

Narrative

Additional Manufacturer Narrative

THE INVESTIGATION OF THE ISSUE IS STILL IN PROGRESS.

Description of Event or Problem

THERE ARE NO REPORTS OF ANY PATIENT OR USER HARM RELATED TO THIS ISSUE. ALIGNRT (AND GATERT) CAN BE INTERFACED TO GATE THE TREATMENT BEAM ON SIEMENS LINACS. WHILST CONNECTED TO A SIEMENS LINAC, FOR REASONS STILL UNDER INVESTIGATION THE ALIGNRT SYSTEM MAY REPORT THE FOLLOWING WARNING MESSAGE DURING A PERIOD OF BEAM-HOLDING DUE TO THE PATIENT BEING OUT OF POSITION: "LINAC BEAM STATUS IS NOT AS SET. HOLD THE BEAM FROM THE LINAC CONSOLE. GATING HAS BEEN DISABLED. RESTART ALIGNRT TO ENABLE GATING". WHILE THE WARNING MESSAGE IS DISPLAYED AND FROM THIS POINT ONWARDS, THE ALIGNRT SYSTEM MAY ALLOW THE BEAM ON THE LINAC TO BE ENABLED INDEPENDENTLY OF THE PATIENT POSITION. THIS OCCURRED ONLY AT ONE CUSTOMER'S SITE. THE CUSTOMER CONFIRMED THAT THE ISSUE WAS PROMPTLY IDENTIFIED AND THAT NO PATIENT HARM OCCURRED. THERE ARE NO REPORTS OF A SIMILAR ISSUE HAVING OCCURRED AT OTHER SITES.