inforMED
InjuryHSH

UNKNOWN DEPUY 3 MBT TRAY

Received Nov 23, 2015 · Event occurred Nov 20, 2015

Report 1818910-2015-35760 · MDR key 5242883

Device

Generic name

Knee Tibial Tray

Catalog number

UNK-KNEE

Lot number

UNK

Product problems

  • Loss of or Failure to Bond

Patient

75 YR

  • No Information

Narrative

Additional Manufacturer Narrative

DEPUY SYNTHES HAS BEEN INFORMED THAT THE CATALOG NUMBER AND LOT NUMBER IS NOT AVAILABLE. THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem

PATIENT WAS REVISED DUE TO A LOOSE TIBIAL TRAY AT THE CEMENT TO IMPLANT INTERFACE. UNKNOWN CEMENT WAS USED.

Additional Manufacturer Narrative

NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE LOT SPECIFIC COMPLAINT DATABASE SEARCH, OR DEVICE HISTORY RECORD (DHR) REVIEW WAS NOT POSSIBLE BECAUSE THE REQUIRED LOT CODE(S) WAS NOT PROVIDED. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.