inforMED
InjuryLBH

MI VARNISH

Received Oct 5, 2015 · Event occurred Sep 3, 2015

Report 1410097-2015-00005 · MDR key 5200530

Device

Generic name

Fluoride Varnish

Manufacturer

Gc Corp.

Catalog number

004274

Lot number

1501071

Product problems

  • Improper or Incorrect Procedure or Method
  • Malposition of Device

Patient

Not reported

  • Burning Sensation
  • Discomfort
  • Sensitivity of Teeth

Narrative

Description of Event or Problem

GC AMERICA RECEIVED A REPORT FROM THE PATIENT'S HYGIENIST ON (B)(6) 2015 VIA PHONE. THE HYGIENIST REPORTED THAT PATIENT EXPERIENCED SENSITIVITY ALONG THE GUM LINE. THE HYGIENIST STATED THAT AFTER THE DISCOMFORT WENT A WAY AS SOON AS THEY RINSED THE VARNISH AWAY. THE HYGIENIST WAS ABLE TO CONFIRM THAT THE PATIENT DID NOT HAVE ANY ALLERGIES. THERE WAS NO ADDITIONAL PATIENT INJURY OR MEDICAL INTERVENTION NEEDED. PATIENT REPORTED BURNING SENSATION ON SOFT TISSUE. HYGIENIST REPORTED APPLYING A THICK LAYER OF VARNISH. THERE IS A POSSIBILITY MI VARNISH CAME IN CONTACT WITH SOFT TISSUE LEADING TO BURNING SENSATION. MI VARNISH IS TO BE APPLIED TO TOOTH ONLY.