inforMED
MalfunctionDPW

BOOD FLOW METER

Received Nov 15, 1996 · Event occurred Aug 23, 1996

Report MW4001715 · MDR key 51967

Device

Generic name

Blood Flow Unit

Model number

TRANSONIC

Catalog number

*

Lot number

*

Product problems

  • Smoking

Patient

82 YR

  • No Consequences Or Impact To Patient

Narrative

Description of Event or Problem

MD USING CAUTERY. FLOW UNIT STARTED TO EMIT SMOKE, SPARKS AND FUMES. UNIT REMOVED. FAILURE APPARENTLY CAUSED BY A "SHORT" IN A CERAMIC CAPACITOR. FLOW METER WAS LEFT PLUGGED IN WHILE CAUTERY IN USE. MFR INSTRUCTIONS STATE TO TURN OFF FLOW METER WHEN NOT IN USE TO PREVENT PROBLEMS OF THIS SORT.