inforMED
MalfunctionKRL

SARNS 8000 PERFUSION SYSTEM

Received Oct 23, 2015 · Event occurred Sep 30, 2015

Report 1828100-2015-00907 · MDR key 5171591

Add to watch list
Product code (KRL)
Model (5773)

Device

Generic name

Detector, Bubble, Cardiopulmonary Bypass-sarns 8000

Model number

5773

Catalog number

5773

Product problems

  • Break

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

(B)(4). THE FIELD SERVICE REPRESENTATIVE (FSR) CONFIRMED THE BROKEN DRAW LATCH. HE REPLACED THE DRAW LATCH ASSEMBLY AND TESTED OPERATION SATISFACTORY. THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. THE SUSPECT PART WAS RETURNED TO THE MANUFACTURER. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem

IT WAS REPORTED THAT DURING SET-UP OF THE DEVICE FOR A CARDIOPULMONARY BYPASS PROCEDURE, THE DRAW LATCH ON ULTRASONIC AIR SENSOR (UAS) BROKE. THE 8K WAS BEING SET-UP FOR A CASE THE NEXT DAY. THEY HAVE A SPARE UAS AND SWAPPED IT FOR THE ONE WITH THE BROKEN DRAW LATCH. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO DELAY, NO BLOOD LOSS, NOR ADVERSE CONSEQUENCES TO THE PATIENT.