inforMED
OtherKPW

EIMAC-VARIAN

Received Nov 19, 1996 · Event occurred Nov 11, 1996

Report 51396 · MDR key 51396

Device

Generic name

Beam Limiting Device

Model number

S 90968

Catalog number

NA

Lot number

NA

Product problems

  • Detachment Of Device Component
  • Component Falling
  • Installation error

Patient

49 YR

  • Bruise/Contusion
  • Headache
  • Pain

Narrative

Description of Event or Problem

PT WAS LYING ON THE X-RAY TABLE HAVING X-RAY TAKEN OF HER MANDIBLE. THE EXTENSION CONE WAS ON THE MACHINE. A SINGLE AP FILM HAD BEEN TAKEN. THE MACHINE WAS BEING MOVED AWAY FROM THE PT TO TAKE ANOTHER VIEW. THE MACHINE WAS ANGLED AND THE CONE FELL FROM THE MACHINE STRIKING THE PT'S HEAD. CONE INSTALLED WRONG IN 1990; SIDE MOUNT INSTEAD OF FRONT MOUNT. NO SPRING LOCK INSTALLED. LAST INSPECTION 5/29/96.