H. PYLORI STAT - PAK

Received Dec 7, 2004 · Event occurred Oct 13, 2004

Report 2431980-2004-00001 · MDR key 5101371

Device

Generic name

Ivd. Rapid Detection Of Ab To H. Pylori

Catalog number

HP133

Lot number

HP111703/1

Product problems

  • False Negative Result

Patient

Not reported

  • No Consequences Or Impact To Patient

Narrative

Additional Manufacturer Narrative

THE PRODUCT LOT IN QUESTION WAS BULK MATERIAL FOR FURTHER PACKAGING BY (B)(4). EVAL OF DEVICE SHOWED THAT IT MET RELEASE CRITERIA.

Description of Event or Problem

ON (B)(6) 2004, A DISTRIBUTOR/REPACKAGER. (B)(4), EMAILED SAYING THAT HP111703/1 OF H. PYLORI STAT- PAK HAS GIVEN 18 NEGATIVES IN A SUSPECT POPULATION. THREE OF THESE WERE CONFIRMED POSITIVE BY ELISA. THERE WAS NO INVOLVEMENT IN ANY DEATH OR SERIOUS INJURY.

Remedial action

  • Replace