VITEK® 2 GRAM-NEGATIVE IDENTIFICATION TEST KIT
Received Sep 10, 2015
Report 1950204-2015-00026 · MDR key 5066992
Device
Product problems
- Incorrect, Inadequate or Imprecise Result or Readings
Patient
Not reported
- No Known Impact Or Consequence To Patient
Narrative
Additional Manufacturer Narrative
DEVICE NOT RETURNED TO MANUFACTURER.
Description of Event or Problem
A CUSTOMER IN (B)(6) RECEIVED CAP RESULTS WHERE THE ISOLATE SHOULD HAVE BEEN IDENTIFIED AS ACINETOBACTER IWOFFII; HOWEVER, RESULTS FROM VITEK 2 GN ID TEST KIT, (REF 21341) WERE IDENTIFIED AS ACINETOBACTER BAUMANII. THE CUSTOMER PERFORMED THE TEST USING TWO DIFFERENT LOTS OF VITEK 2 GN ID TEST KIT (LOTS 241337120, MANUF. 08MAR15, EXPIRY 07MAR16 AND 241344640, MANUF. 22MAY15, EXPIRY 21MAY16), AND OBTAINED THE SAME RESULTS. THERE IS NO INDICATION OR REPORT FROM THE HOSPITAL OR TREATING PHYSICIAN TO BIOMERIEUX THAT THE ORGANISM MISIDENTIFICATION LED TO ANY ADVERSE EVENT RELATED TO THE PATIENT'S STATE OF HEALTH. LAB REPORTS AND ISOLATE SUBMITTAL HAVE BEEN REQUESTED FROM THE CUSTOMER. BIOMERIEUX INVESTIGATION HAS BEEN INITIATED.
Additional Manufacturer Narrative
BIOMÉRIEUX INVESTIGATION WAS CONDUCTED. THE ORGANISM WAS SUBCULTURED, AND TESTING INCLUDED ONE (1) GN ID CARD FROM EACH CUSTOMER LOT AND ONE (1) CARD FROM TWO (2) RANDOM LOTS. API 20NE WAS ALSO PERFORMED. FOR ONE (1) OF THE CUSTOMER LOTS (241337120) AND THE TWO (2) RANDOM LOTS, A LOW DISCRIMINATION CALL OF ACINETOBACTER LWOFFII / MORAXELLA GROUP WAS OBTAINED. SINCE ACINETOBACTER LWOFFII IS PRESENTED AS A POSSIBLE ORGANISM, THE RESULTS ARE ACCEPTABLE FOR VITEK 2 TESTING. FOR THE OTHER CUSTOMER LOT (241344640), AN IDENTIFICATION OF MORAXELLA GROUP WAS OBTAINED. THE API 20NE GAVE A GOOD IDENTIFICATION (B)(4)OF A. LWOFFII . DUE TO THE DUPLICATION OF THE CUSTOMER RESULT FOR GN ID CARD LOT 241344640, EVALUATION OF MANUFACTURING AND QC BATCH RECORDS WAS PERFORMED. THE EVALUATION INDICATED THE LOT PASSED ON INITIAL TESTING AND PERFORMED IN ACCORDANCE WITH SPECIFICATIONS.