WARMER DRAPE
Received Jan 30, 2015 · Event occurred Dec 12, 2014
Report 8043817-2015-00001 · MDR key 5046368
Device
Generic name
Fluid Warming And Slush Drapes
Manufacturer
Microtek Dominicana, S.a.Model number
ORS-100Catalog number
ORS-100
Lot number
D133431
Product problems
- Fluid/Blood Leak
Patient
45 YR
- No Code Available
Narrative
Additional Manufacturer Narrative
AN INVESTIGATION IS COMPLETED AT THE MFG SITE IN THE (B)(4). BASED ON THE DHR REVIEWED, THIS LOT HAD (B)(4) PIECES AND IT WAS MANUFACTURED ON 12/09/2013. NO DEFECTS WERE REPORTED DURING PROCESSING, PACKAGING OR FINAL INSPECTION. THE INVESTIGATION REVEALED THAT THIS PROBLEM WAS NOT DUE TO AN ASSEMBLY, MATERIAL, OR PROCESS ISSUE. (B)(4)'S COMPLAINT DEPT. HAS CONTACTED (EMAIL AND CALLED) THE USER FACILITY, (B)(6), THREE TIMES PER THE FOLLOWING DATES: (B)(4) 2015. THERE HAS BEEN NO RESPONSE FROM THE USER FACILITY TO THE (B)(4) COMPLAINT DEPT. NO PRODUCT HAS BEEN RETURNED. SINCE NO DEVICE WAS RETURNED FOR ANALYSIS, THE ROOT CAUSE FOR THE EVENT HAS NOT BEEN DETERMINED.
Description of Event or Problem
THERE WAS A REPORT FORM THE USER FACILITY (UF/IMPORTER REPORT#: (B)(4)), (B)(6). THAT REPORT WAS SENT TO (B)(4); RECEIVED (B)(4) 2014. PRODUCT#: ORS-100, WARMER DRAPES, HAD BEEN FOUND LEAKING. UPON TAKING THE IRRIGATION DRAPE OUT OF THE FLUID WARMER AT THE END OF PROCEDURE, A PIN HOLE WAS NOTED IN THE DRAPE. SALINE IRRIGATION WAS LEAKING THROUGH. THE NURSE ANESTHETIST (CRNA) RE-DOSED THE PATIENT WITH 1 GRAM OF CEFAZOLIN AT THE END OF THE PROCEDURE. L KNEE ACL RECONSTRUCTION WAS THE INTENDED PROCEDURE.